Prof. Dr. Thomas Baier / DG8SAQ Juli 2006 Prof. Dr. Thomas Baier DG8SAQ Prof. Dr. Thomas Baier / DG8SAQ Juli 2006 Minimum frequency range 100 KHz … 30 MHz Sufficient accuracy to use S-parameters for simulations Utilize standard PC as much as possible Cut down on hardware as much as possible Easily obtainable standard components, solderable by amateur with soldering iron
Transcript
Page 1: Prof. Dr. Thomas Baier DG8SAQsdr-kits.net/DG8SAQ/VNWA/VNWA_English.pdf · Prof. Dr. Thomas Baier / DG8SAQ Juli 2006 VNWA HP8753C Verbesserung durch 10 dB Dämpfungsglied vor RX-Eingang

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

Prof. Dr. Thomas BaierDG8SAQ

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

• Minimum frequency range 100 KHz … 30 MHz

• Sufficient accuracy to useS-parameters for simulations

• Utilize standard PC as much as possible • Cut down on hardware as much as

possible• Easily obtainable standard components,

solderable by amateur with soldering iron

Page 2: Prof. Dr. Thomas Baier DG8SAQsdr-kits.net/DG8SAQ/VNWA/VNWA_English.pdf · Prof. Dr. Thomas Baier / DG8SAQ Juli 2006 VNWA HP8753C Verbesserung durch 10 dB Dämpfungsglied vor RX-Eingang

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

Page 3: Prof. Dr. Thomas Baier DG8SAQsdr-kits.net/DG8SAQ/VNWA/VNWA_English.pdf · Prof. Dr. Thomas Baier / DG8SAQ Juli 2006 VNWA HP8753C Verbesserung durch 10 dB Dämpfungsglied vor RX-Eingang

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

from datasheet AD9851, Analog Devices

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

Computational Example: clock frequency RF-DDS: 180 MHzclock frequency LO-DDS: 170 MHz IF: 10 MHz

DDS Alias Frequencies

0

50

100

150

200

250

300

350

400

0 30 60 90 120 150 180 210 240 270 300 330 360 390

RF-Frequeny

Spec

ta o

f RF

and

LO D

DS

Page 4: Prof. Dr. Thomas Baier DG8SAQsdr-kits.net/DG8SAQ/VNWA/VNWA_English.pdf · Prof. Dr. Thomas Baier / DG8SAQ Juli 2006 VNWA HP8753C Verbesserung durch 10 dB Dämpfungsglied vor RX-Eingang

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

≈ 30 kHz

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

≈ 30 kHz

Usable frequency range: 200 Hz … about 500 MHzExceptions:Zero signal amplitude in vicinitiy of integer multiples of DDS clock frequency (180MHz, 360 MHz, 540 MHz …) → measurements not possible

Page 5: Prof. Dr. Thomas Baier DG8SAQsdr-kits.net/DG8SAQ/VNWA/VNWA_English.pdf · Prof. Dr. Thomas Baier / DG8SAQ Juli 2006 VNWA HP8753C Verbesserung durch 10 dB Dämpfungsglied vor RX-Eingang

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

Problem: make wideband BALUNSolution: use balanced Gilbert Cell Mixer

BALUN

RF

1 : 1

50

50

50

DUT

Bridge Voltage

50

50

ZS = 50

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

RF+ RF-

IF+ IF -

LO

Page 6: Prof. Dr. Thomas Baier DG8SAQsdr-kits.net/DG8SAQ/VNWA/VNWA_English.pdf · Prof. Dr. Thomas Baier / DG8SAQ Juli 2006 VNWA HP8753C Verbesserung durch 10 dB Dämpfungsglied vor RX-Eingang

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

Problem: Frequenzgang des DC-Blocks im Hz - Bereich

RF DDSZS = ∞

8,2

27

15

DUT

BridgeVoltage

10

ZS = 50,2

LO DDS

Reference Signal= Bridge Current

LO DDS

+

-

+

-

NE612

NE612

10 F Tantal

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

TX RX

to Parallel Interface to Line In +12V/250mA DC

100mm

Page 7: Prof. Dr. Thomas Baier DG8SAQsdr-kits.net/DG8SAQ/VNWA/VNWA_English.pdf · Prof. Dr. Thomas Baier / DG8SAQ Juli 2006 VNWA HP8753C Verbesserung durch 10 dB Dämpfungsglied vor RX-Eingang

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

Page 8: Prof. Dr. Thomas Baier DG8SAQsdr-kits.net/DG8SAQ/VNWA/VNWA_English.pdf · Prof. Dr. Thomas Baier / DG8SAQ Juli 2006 VNWA HP8753C Verbesserung durch 10 dB Dämpfungsglied vor RX-Eingang

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

LOAD "F64A.s1p"

LOAD "F64B.s1p"

Circuit Diagram [...] (E)Sweep Init (E)

Sweep "Initial analysis"LOOP 400 FREQ LIN 10.685Meg 10.714925MegW 0 Y "Gain" "dB" -30 0W 1 SMITH

Show + W 0 Y MagdB(S(2,1)) + W 0 Y MagdB(S(1,1)) + W 0 Y MagdB(S(2,2)) + W 1 DB S(1,1) + DB S(2,2)

EndSweep

port_2 270ohmport_1 270ohm

… e.g. with APLAC:

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

10.685M 10.692M 10.700M 10.707M 10.715M

-30.00

-22.50

-15.00

-7.50

0.00

Initial analysisAPLAC 8.00 Student version FOR NON-COMMERCIAL USE ONLY

Gain

dB

f/HzMagdB(S(2,1)) MagdB(S(1,1))MagdB(S(2,2))

Page 9: Prof. Dr. Thomas Baier DG8SAQsdr-kits.net/DG8SAQ/VNWA/VNWA_English.pdf · Prof. Dr. Thomas Baier / DG8SAQ Juli 2006 VNWA HP8753C Verbesserung durch 10 dB Dämpfungsglied vor RX-Eingang

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

VNWAZS = ZL = 50 Ω

VNWAsoftwarematched

HP8753Csoftwarematched

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

simulated matchingresistive matching

ZS = ZL = 50 Ω

ZS = ZL = 610 Ω

e.g. 11 kHz LC Filter

Page 10: Prof. Dr. Thomas Baier DG8SAQsdr-kits.net/DG8SAQ/VNWA/VNWA_English.pdf · Prof. Dr. Thomas Baier / DG8SAQ Juli 2006 VNWA HP8753C Verbesserung durch 10 dB Dämpfungsglied vor RX-Eingang

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

VNWA

HP8753C

Verbesserung durch 10 dB Dämpfungsglied vor RX-Eingang

e.g. 400 MHz SAW Filter

Prof. Dr. Thomas Baier / DG8SAQ Juli 2006

Achieved:• 200 Hz – 500 MHz covered with exceptions,

especially VLF-2m, 70cm covered!• Accuracy sufficient for network synthesis.• Minimum hardware, no tuning points.• Universal software• Principally achievable accuracy with PC sound

device: 0,01dB and 0,1°,currently limited by Mixer nonlinearities

Possible Improvements:• Higher level mixers• DDS devices with higher clock rates (up to 1 GHz)